background image

4-2 Basic Settings and Advanced Settings

4-11

IB-E

4

Set
ting V

a

rious

F

unct

ions

Set the averaging or high-pass filter for the data retrieved at the sampling rate.

Averaging

The average value is moved and averaged. If the measured values fluctuate, stable
measurements can be obtained by increasing the average number of times. When [HPF] is
set, the high-pass filter will be enabled.

List of averaging

The sampling rate, R.V. update cycle and response time differ according to the auto adjust
function settings.

Auto adjust function disabled

3. Averaging/High-Pass Filter

Item

Setting range

Default value

Averaging

1

/2/4/8/16/64/256/1024/4096/16384/HPF

64

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

#X'

GO

HOLD

CALC

CHECK

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

#X'

GO

HOLD

CALC

CHECK

PQ

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

QWV

GO

HOLD

CALC

CHECK

3. Averaging/

High-pass filter

Averaging setting
screen

4. Output mode

/ button

[MODE]
or 

(page 4-11)

(page 4-11)

(page 4-13)

Averaging

Sampling rate

(ms)

R.V. refresh rate

(ms)

Response

time (ms)

1

0.08

0.08

0.16

2

0.24

4

0.4

8

0.72

16

1.36

64

5.2

256

20.56

1024

0.32

82.24

4096

1.28

328.96

16384

5.12

1315.84