Setting Delay Volume

Installation Guide

| 95

tubing length between the UV outlet port and outlet valve as well as the tubing
length between the outlet valve and the fraction collector drophead.

8.

Select or clear the next four checkboxes depending on whether or not these
modules are present on your system.

Note:

If a pH detector module is present on your system and you select

the checkbox to include its cell volume, determine whether the detector is

set to Inline or Bypass mode and select the appropriate radio button.

9.

If one or more of the options is not applicable or if you change the length of the
tubing, enter the appropriate values in the Additional volume box.

The system calculates the delay volume based on your specifications and

displays the results in

μ

l at the bottom of the dialog box.

10. Click OK.