Ngc system location is a standard cold cabinet – Bio-Rad ChromLab™ Software User Manual
Page 36
3
| Setting Up NGC Chromatography Systems
34
| NGC Chromatography Systems and ChromLab Software
NGC System Location Is a Standard Cold Cabinet
To connect the touch screen to the NGC
system when the system is located in a
cold cabinet perform the following tasks:
Connect the touch screen and communication cables to the NGC
instrument before inserting it into the cold cabinet
Place the instrument in the cold cabinet
Connect the touch screen to the touch screen stand
Connect the touch screen cable to the touch screen
These tasks are explained in detail in the sections that follow.
To connect the touch screen and communication cables
1.
Determine the side of the cold cabinet into which you plan to install the
instrument and place a table on that side of the cabinet.
2.
Locate the touch screen display cable in the accessories kit or the NGC
touch
screen bracket kit (#788-5060).
Tip:
Bio-Rad provides a standard 5' (1.52 m) cable in the accessories kit
and a 15' cable with the NGC
touch screen bracket kit or the NGC touch
screen stand (#788-5040). A 10' (3.0 m) cable is also available.
3.
Insert the touch screen cable into the touch screen connector port on the back
of the instrument and tighten the thumbscrews.