Plumbing NGC Systems

Installation Guide

| 103

8.

Attach a 30 ml syringe to the 1/4-28 female to male luer adaptor.

9.

Withdraw the plunger on the syringe until liquid is visible in the syringe.

10. Detach the luer adaptor and syringe from the pump head wash priming tubing

on pump B.

11. Locate the second inlet tubing with the 1/4-28 union that you set aside in

step 7.

12. Connect the union to the pump head wash priming tubing on system pump B.

13. Immerse the other end of the inlet tubing in the tube of 20% IPA.

To plumb the sample pump

Perform the steps in the section

To plumb and prime the system pump wash lines

on page

102

with the following exception:

Attach the inlet tubing to the left

rear port on the sample pump, and the pump

head wash priming tubing to the right rear port.