Inserting Additional Modules

Installation Guide

| 51

Inserting Additional Modules

WARNING!

Disconnect power to the NGC instrument before inserting or

removing any module. No user-serviceable parts are inside any component

unless noted in this manual. Refer servicing questions to Bio-Rad service

personnel.

WARNING!

To reduce the chance of liquid seeping into the instrument, all

open bays must be filled with at least an empty module. The NGC instrument

will not operate if it detects a slot without a module inserted.

To insert or reposition modules on the NGC instrument

1.

Ensure that the NGC instrument is powered off.

2.

Loosen the screws on the front of the module cover or covers.

3.

Firmly grasp both captive screws and pull the module forward out of its bay.

The following image shows an empty bay.

4.

Store the module and the screws in a safe place for future use.