ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 90

background image

C

HAPTER

6 M

ONITORING

ECG

6-4

www.zoll.com

9650-001355-01 Rev. H

5-Lead Electrode Placement

Depending upon local usage, the ECG leads are marked either RA, LA, LL, RL, and V or R, L,
F, N and C. The following table shows the markings and color codes for the different lead sets.

AHA Color Coding

IEC Color Coding

Placement of Electrodes

RA/White Electrode

R/Red Electrode

Place near patient’s right mid-clavicular line,
directly below clavicle.

LA/Black Electrode

L/Yellow Electrode

Place near patient’s left mid-clavicular line,
directly below clavicle.

LL/Red Electrode

F/Green Electrode

Place between 6th and 7th intercostal
space on patient’s left mid-clavicular line.

RL/Green* Electrode

N/Black* Electrode

Place between 6th and 7th intercostal
space on patient’s right mid-clavicular line.

V/Brown* Electrode

C/White* Electrode

Single movable chest electrode. Place this
electrode in one of the positions, V1 - V6,
as shown in the following figure.

V1 -- 4th intercostal space at right sternal
margin.

V2 -- 4th intercostal space at left sternal
margin.

V3 -- Midway between V2 and V4 leads.

V4 -- 5th intercostal space at mid-clavicular
line.

V5 -- Same transverse level as V4 at left
anterior-axillary line.

V6 -- Same transverse level as V4 at left
mid-axillary line.