Pulse oximeter – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 306
Specifications
A-20
www.zoll.com
9650-001355-01 Rev. H
Respiration Rate Accuracy:
0 to 70 bpm:
±1 bpm
71 to 120 bpm: ±2 bpm
121 to 149 bpm: ±3 bpm
Flow rate: 50 ml/min -7.5 + 15 ml/min, flow measured by volume.
Total System Response Time:
2.9 seconds typical, 3.9 seconds maximum.
Pulse Oximeter
Range:
Oxygen Saturation (% SpO
2
)
1% - 100%
Carboxyhemoglobin Saturation (% SpCO)
0% - 99%
Methemoglobin Saturation (% SpMet)
0% - 99%
Pulse Rate (bpm)
25 - 240 beats per minute
Accuracy:
Oxygen Saturation (% SpO
2
) - During No Motion Conditions
Adults, Pediatrics
1
70% - 100%, ±2 digits 0% - 69%, unspecified
Neonates
2
70% - 100%, ±3 digits 0% - 69%, unspecified
Oxygen Saturation (% SpO
2
) - During Motion Conditions
3
Adults, Pediatrics
70% - 100%, ±3 digits 0% - 69%, unspecified
Neonates
70% - 100%, ±3 digits 0% - 69%, unspecified
Oxygen Saturation (% SpO
2
) - During Low Perfusion Conditions
4
Adults, Pediatrics
70% - 100%, ±2 digits
Neonates
70% - 100%, ±3 digits
Carboxyhemoglobin Saturation (% SpCO)
5
1% - 40% ±3 digits
Methemoglobin Saturation (% SpMet)
5
1% - 15% ±1 digit
Pulse Rate (bpm) - During No Motion Conditions
1
Adults, Pediatrics, Neonates
25 - 240 ±3 digits
Pulse Rate (bpm) - During Motion Conditions
3
Adults, Pediatrics, Neonates
25 - 240 ±5 digits
Resolution: SpO
2
: 1%
SpCO: 1%
SpMet: 0.1% for range up to 9.9%, 1% for range 10 - 99%
Pulse rate: 1 bpm (beats per minute)
Alarm Limits: On/Off displayed on monitor. User selectable.
SpO
2
: High 72 - 100% saturation, Low 70 - 98% saturation
SpCO: High 2 - 100% saturation, Low 0 - 99% saturation
SpMet: High 1 - 100% saturation, Low 0 - 99% saturation
Pulse Rate: High 60 - 235 beats per minute, Low 20 - 100 beats per minute