Ibp setup, Attaching the invasive pressure transducer – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 150
C
HAPTER
11 M
ONITORING
I
NVASIVE
P
RESSURES
(IBP)
11-2
www.zoll.com
9650-001355-01 Rev. H
Warning!
•
If electrocautery is used, always avoid using any transducer with a conductive
(metal) case connected to its ground shield. Using a conductive transducer case that
is connected to its cable shield risks high-frequency burns at the ECG electrodes if
the transducer case becomes grounded to earth.
•
Normal alarm functions will detect complete disconnections of invasive pressure
transducers; however, the alarm functions will not detect a partial disconnection or
the use of some incompatible transducers. Use only approved transducers and ensure
that the transducers are connected properly.
•
Before you use the X Series unit on a new patient, always turn it off for at least 2
minutes. This clears the previous patient’s trend values, alarm limit settings, and
NIBP cuff inflation pressure.
•
Use only ZOLL-approved IBP sensors. Use of non-approved sensors may result in
inaccurate IBP measurements.
IBP Setup
To take safe and accurate IBP measurements using the X Series, you must perform the
following steps, each of which corresponds to a section in this chapter. Read each section
carefully before you perform IBP measurements.
1. Attach the invasive pressure transducer to the X Series unit.
2.
Zero the transducer.
3.
Set the invasive pressure alarms (according to your organization’s standards).
4.
Select a label for the invasive pressure channel.
Attaching the Invasive Pressure Transducer
Follow these steps when attaching the invasive pressure transducer:
1. Inspect the transducer cable. If the cable shows any signs of wear, breakage, or fraying,
replace the cable. Replace the transducer dome, if necessary.
2.
Apply the transducer according to your organization’s procedures. Always refer the
manufacturer’s Directions for Use before using a transducer.
3.
If the transducer is a disposable unit with a separate cable, connect the transducer to the
transducer cable.
4.
Plug the transducer cable into one of the three six-pin IBP cable connectors on the side of the
X Series unit.