Using the co2 parameter control panel – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 132
C
HAPTER
9 M
ONITORING
CO2
9-10
www.zoll.com
9650-001355-01 Rev. H
Setting Upper and Lower Respiratory Rate (RR/BR) Limits
Initially, the RR/BR Alarm menu specifies that RR/BR alarms are enabled (
ON
) or disabled
(
OFF
), and displays the default Upper and Lower respiratory limits. The upper and lower limits
can be
ON
or
OFF
(default is
OFF
). The following table lists the default Respiratory limits for
adult, pediatric, and neonate patients in Breaths/Minute, and gives the range in which you can
set these limits:
Using the CO2 Parameter Control Panel
Use the navigation keys to highlight and select the CO2 numeric display to display the
CO2 Parameter Control Panel:
Figure 9-2 CO2 Parameter Control Menu
On the CO2 Parameter Control Panel, you can select a CO
2
alarm (EtCO2, Breath Rate, or
FiCO2). On the selected alarm settings menu, you can enable/disable alarms and set alarm
limits (as described previously). You can also set the CO
2
sweep speed.
Patient Type
Respiration Rate Default
Respiration Rate Range
Adult
Lower: 3 BPM
Upper: 50 BPM
Lower: 0 to 145 BPM
Upper: 5 to 150 BPM
Pediatric
Lower: 3 BPM
Upper: 50 BPM
Lower: 0 to 145 BPM
Upper: 5 to 150 BPM
Neonate
Lower: 12 BPM
Upper: 80 BPM
Lower: 0 to 145 BPM
Upper: 5 to 150 BPM
0 to 30 mmHg
0 to 150 mmHg
CO2 Scale
0 to 100 mmHg
0 to 60 mmHg
CO2
Lower
Upper
EtCO2 Alarm
28
50
No Breath Alarm
FiCO2 Alarm
0
10
30 sec
30
RR/BR Alarm
10
CO2/Resp Sweep Speed
12.5 mm/s