Recorder

9650-001355-01 Rev. H

X Series Operator’s Guide

A-17

Recorder

Type: High-resolution thermal array.

Annotation: Time, date, ECG lead, ECG gain, heart rate, defibrillation and pacing parameters
and treatment summary events.

Paper Width: 80 mm.

Paper Speed: 25 mm/sec, 50 mm/sec

Delay: 6 seconds.

Frequency Response: Automatically set to monitor's frequency response.

Treatment Summary:

10 switches to record key ACLS events (IV, INTUB, EPI, LIDO, ATROP, etc.).
Automatically logs into memory the type of event, time and ECG sample.

Full Disclosure Case Log: A combination of 32 monitor snapshots; 500 non-ECG events; and
24 hours of continuous ECG (4 waveforms), Capnography, IBP (3 channels), and Pads
Impedance. The actual information stored can be more or less depending on the use profile and
the log configuration settings.

Record Modes: Manual and automatic (User-configurable).

Battery

Type: Rechargeable Lithium-Ion, 11.1Vdc, 6.6 Ah, 73Wh

Capacity:

With a new, fully charged battery operating at room temperature:

At least

6 hours of continuous monitoring of ECG, SpO

2

, CO

2

, three Invasive Pressure

channels, and 2 channels of Temperature, with NIBP measurements every 15 minutes and
10 200 J shocks (display set to 30%).

At least 100 discharges at maximum shock energy (200 joules).

At least 3.5 hours pacing, with ECG, SPO

2

, CO

2

, three Invasive Pressures, temperature,

NIBP every 15 minutes and pacing at 180 ppm, and 140 mA.

At least 10 discharges at maximum shock setting (200 joules) after a Low Battery
indication.

Note:

Proper battery care is required to maintain maximum available capacity.

Battery Indicators:

5 Battery capacity LED indicators, Fault indicator, Recalibration indicator

Recharge Rate: 100% in 4 hours, when initiated at Low Battery indication.