beautypg.com

Begin cpr following local medical protocols, Turn on unit, 1 select energy level – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 192

background image

C

HAPTER

15

M

ANUAL

D

EFIBRILLATION

15–2

www.zoll.com

9650-001355-01 Rev. H

Determine the Patient’s Condition Following Local Medical Protocols

Verify the following:

Unconsciousness

Absence of breathing

Absence of pulse

Begin CPR Following Local Medical Protocols

Request additional assistance.

Turn On Unit

Press the green power switch located on the top of the unit. The green, yellow, and red lights at
the top of the unit flash on and off, and the unit displays the message SELF TEST PASSED.

1 Select Energy Level

Press the Energy Select arrows up or down to select the desired energy level. These buttons are
located either on the front of the unit or on the STERNUM paddle.

Note:

The initial defibrillator pad/external paddle energy selection can be set in Defib/Pacer
default settings in the Setup>Supervisor menu. The energy selections listed for
Shock 1, Shock 2, and Shock 3 are used only when Basic Auto Energy Escalation is
enabled. Otherwise, the values set in Defib/Pacer > Default Settings are used.

The default energy selections for adult patients are:

Shock 1 - 120 joules
Shock 2 - 150 joules
Shock 3 - 200 joules

The default energy selections for pediatric patients are:

Shock 1 - 50 joules
Shock 2 - 70 joules
Shock 3 - 85 joules

Note:

Neonatal and pediatric defibrillator energy levels should be selected based on site-
specific protocols.

The selected energy level is displayed at the bottom of the display screen.

SHOCK

1

2

3

or