beautypg.com

Begin cpr following medical protocols, Prepare patient – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 196

background image

C

HAPTER

15

M

ANUAL

D

EFIBRILLATION

15–6

www.zoll.com

9650-001355-01 Rev. H

Emergency Defibrillation Procedure with Hands-Free
Therapy Electrodes

Determine the Patient’s Condition Following Local Medical Protocols

Verify:

Unconsciousness

Absence of breathing

Absence of pulse

Begin CPR Following Medical Protocols

Request additional assistance.

Prepare Patient

Remove all clothing covering the patient’s chest. Dry chest if necessary. If the patient has
excessive chest hair, clip or shave it to ensure proper adhesion of the electrodes.

Attach hands-free therapy electrodes according to instructions on the electrode packaging.

Ensure that the therapy electrodes are making good contact with the patient’s skin and are not
covering any part of the ECG electrodes.

Therapy Electrode Application

Warning!

Poor adherence and/or air under the therapy electrodes can lead to the possibility of
arcing and skin burns.

1. Apply one edge of the pad securely to the patient.

2.

Roll the pad smoothly from the applied edge to the other, being careful not to trap any air
pockets between the gel and skin.

ZOLL hands-free therapy electrodes are a defibrillation-protected Type BF patient
connection.

ECG leads are a defibrillation-protected Type CF patient connection.