- ZOLL Base PowerCharger1x1 Operator’s Manual
- Section 1
- General Information
- Warnings and Cautions
- Section 2
- Operation
- 1. Check to see that the unit is connected to live AC mains. The POWER light should be lit.
- 2. Insert battery into the compartment. Make sure the battery is fully seated (A click can be hea...
- 3. The CHARGER ON and TESTING indicators illuminate and remain on during complete AutoTest and re...
- 4. In 8 hours or less for the PD 4410/Smart Battery or 15 hours or less for the XL Smart Battery ...
- 5. To remove the battery, insert a finger into the recess at the top of the battery compartment, ...
- 1. Check to see that the Base PowerCharger 1x1 is connected to live AC mains. The POWER light sho...
- 2. Insert battery into the charger. Make sure the battery is fully seated. (A click can be heard).
- 3. Press the TEST button. The TESTING indicator light will illuminate for the duration of the tes...
- 4. If the battery passes the test, the BATT. READY light will illuminate while the TESTING light ...
- 5. If a battery fails the test, the FAULT light will illuminate while the TESTING light remains l...
- 6. Record the date and results of the test on the battery label.
- CHARGER ON
- BATT. READY
- FAULT
- TESTING
- Description
- AutoTest with Battery in Compartment
- Manual Test with Battery in Compartment and “TEST” button pushed
- CHARGER ON
- BATT. READY
- FAULT
- TESTING
- Description
- AutoTest with Battery in Compartment
- Manual Test with Battery in Compartment and “TEST” button pushed
- CHARGER ON
- BATT. READY
- FAULT
- TESTING
- Description
- QuickCharge with Battery in Compartment
- Manual Test with Battery in Compartment and “TEST” button pushed
- CHARGER ON
- BATT. READY
- FAULT
- TESTING
- Description
- QuickCharge with Battery in Compartment
- Manual Test with Battery in Compartment and “TEST” button pushed
- Section 3
- Maintenance and Troubleshooting
- 1. Connect the unit to live AC mains. All four indicator lights will illuminate for approximately...
- 2. With the battery compartment empty, check that all indicator lights except POWER are off.
- 3. Press the TEST button to check that the unit is fully operational.
- 4. All four indicator lights will illuminate for approximately 4- 5 seconds.
- 5. If the TESTING light and BATT. READY light remain lit for three seconds while the other lights...
- 6. If the FAULT light remains lit, the unit is defective.
- 7. Insert a battery into the compartment. Verify that the CHARGER ON light illuminates immediately.
- 1. POWER light does not illuminate.
- 2. CHARGER ON indicator light does not illuminate when battery is in compartment.
- 3. TESTING light does not illuminate when TEST button is pressed.
- Section 4
- Battery Management
- EMC Specifications
- Note 1: At 80MHz and 800MHz the higher frequency range applies.
- Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected...
- Note 1: At 80MHz and 800MHz the higher frequency range applies.
- Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected...