Defibrillator – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 288
Specifications
A-2
www.zoll.com
9650-001355-01 Rev. H
Defibrillator
Charge Time:
•
Less than 7 seconds with a new, fully charged battery (first 15 charges to 200 joules).
•
For the sixteenth discharge at maximum energy, the charge time is less than 10 seconds.
Depleted batteries result in a longer defibrillator charge time.
•
Less than 15 seconds when operating without a battery, using AC power alone at 90% of the
rated mains voltage.
•
Less than 25 seconds from the initial power on, with a new, fully charged battery pack
(depleted by up to fifteen 200 joule discharges) or when operating without a battery, using
AC power alone at 90% of the rated mains voltage.
Rhythm Analysis and Charge Time in AED Mode
•
Less than 30 seconds with a new, fully charged battery (first 15 charges to 200 joules).
•
For the sixteenth discharge at maximum energy, the analysis and charge time is less than 30
seconds. Depleted batteries result in a longer defibrillator charge time.
•
Less than 30 seconds when operating without a battery, using AC power alone at 90% of the
rated mains voltage.
•
Less than 40 seconds from the initial power on, with a new, fully charged battery pack
(depleted by up to fifteen 200 joule discharges) or when operating without a battery, using
AC power alone at 90% of the rated mains voltage.
Patient Impedance Range: 10- 300 ohms
Synchronized Mode: Synchronizes defibrillator discharge to the patient’s R wave. SYNC is
indicated on the display with R wave markers above the ECG waveform on the screen and
stripchart. When ECG is monitored by the device, meets the DF-80:2003 requirement of 60ms
maximum time delay between the peak of the R wave and the delivery of energy.
Table A-1 shows the characteristics of the X Series Rectilinear Biphasic™ waveform when
discharged into 25 ohm, 50 ohm, 100 ohm, 125 ohm, 150 ohm and 175 ohm loads at the
maximum energy setting of 200 joules.
Table A-1.
X Series Rectilinear Biphasic Waveform Characteristics
200 J discharged into
25
50
100
125
150
175
First phase
Maximum initial current
31.4 A
30.4 A
19.7 A
19.4 A
16.7 A
15.6 A
Average current
27.1 A
24.9 A
17.5 A
16.2 A
14.4 A
13.2 A
Duration
6 ms
6 ms
6 ms
6 ms
6 ms
6 ms
Interphase duration
(between first and
second phases)
200
s
200
s
200
s
200
s
200
s
200
s
Second phase
Initial current
29.2 A
18.8 A
15.1 A
13.2 A
12.1 A
11 A
Average current
14.7 A
13 A
12.5 A
11.3 A
10.7 A
9.9 A
Duration
4 ms
4 ms
4 ms
4 ms
4 ms
4 ms