Defibrillator

Charge Time:

Less than 7 seconds with a new, fully charged battery (first 15 charges to 200 joules).

For the sixteenth discharge at maximum energy, the charge time is less than 10 seconds.
Depleted batteries result in a longer defibrillator charge time.

Less than 15 seconds when operating without a battery, using AC power alone at 90% of the
rated mains voltage.

Less than 25 seconds from the initial power on, with a new, fully charged battery pack
(depleted by up to fifteen 200 joule discharges) or when operating without a battery, using
AC power alone at 90% of the rated mains voltage.

Rhythm Analysis and Charge Time in AED Mode

Less than 30 seconds with a new, fully charged battery (first 15 charges to 200 joules).

For the sixteenth discharge at maximum energy, the analysis and charge time is less than 30
seconds. Depleted batteries result in a longer defibrillator charge time.

Less than 30 seconds when operating without a battery, using AC power alone at 90% of the
rated mains voltage.

Less than 40 seconds from the initial power on, with a new, fully charged battery pack
(depleted by up to fifteen 200 joule discharges) or when operating without a battery, using
AC power alone at 90% of the rated mains voltage.

Patient Impedance Range: 10- 300 ohms

Synchronized Mode: Synchronizes defibrillator discharge to the patient’s R wave. SYNC is
indicated on the display with R wave markers above the ECG waveform on the screen and
stripchart. When ECG is monitored by the device, meets the DF-80:2003 requirement of 60ms
maximum time delay between the peak of the R wave and the delivery of energy.

Table A-1 shows the characteristics of the X Series Rectilinear Biphasic™ waveform when
discharged into 25 ohm, 50 ohm, 100 ohm, 125 ohm, 150 ohm and 175 ohm loads at the
maximum energy setting of 200 joules.

Table A-1.

X Series Rectilinear Biphasic Waveform Characteristics

200 J discharged into

25

50

100

125

150

175

First phase

Maximum initial current

31.4 A

30.4 A

19.7 A

19.4 A

16.7 A

15.6 A

Average current

27.1 A

24.9 A

17.5 A

16.2 A

14.4 A

13.2 A

Duration

6 ms

6 ms

6 ms

6 ms

6 ms

6 ms

Interphase duration
(between first and
second phases)

200

s

200

s

200

s

200

s

200

s

200

s

Second phase

Initial current

29.2 A

18.8 A

15.1 A

13.2 A

12.1 A

11 A

Average current

14.7 A

13 A

12.5 A

11.3 A

10.7 A

9.9 A

Duration

4 ms

4 ms

4 ms

4 ms

4 ms

4 ms