15 m – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 194

C
HAPTER
15
M
ANUAL
D
EFIBRILLATION
15–4
www.zoll.com
9650-001355-01 Rev. H
To increase or decrease the selected energy after you have pressed the CHARGE button, use 
the defibrillator Energy Select buttons on either the STERNUM paddle or the defibrillator 
front panel.
Caution
Changing the selected energy while the unit is charging or charged causes the defibrillator to 
disarm itself. Press the CHARGE button again to charge the unit to the newly selected energy 
level.
Press the CHARGE button again to confirm the charge.
A charging message displays at the bottom of the screen, and a distinctive charging tone sounds 
indicating that the unit is charging. 
The energy range bar graph on the right side of the display highlights the charge level until it 
reaches the selected energy. When the unit is fully charged, the tone changes to a continuous 
charge ready tone, the highlighted energy bar graph includes the selected energy, and the charge 
indicator on the APEX paddle lights up.
06/06/2011 12:34:56
Adult
00:17:43
I, II,
III...
IBP
12
CO
2
1 cm/mV
II
1
Defibrillator
Biphasic
Confirm Energy
Press CHARGE
1
200
J
120
J
Selected Energy
1
2
4
5
6
8
9
10
15
20
30
50
120
70
150
200
3
7
85
100
Disarm
SYNC
06/06/2011 12:34:56
Adult
00:17:43
I, II,
III...
IBP
12
CO
2
1 cm/mV
II
1
Defibrillator
Biphasic
CHARGED
Press to Shock
1
200
J
120
J
Selected Energy
1
2
4
5
6
8
9
10
15
20
30
50
120
70
150
200
3
7
85
100
Disarm
SYNC