beautypg.com

ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 26

background image

C

HAPTER

1 G

ENERAL

I

NFORMATION

1-14

www.zoll.com

9650-001355-01 Rev. H

Flashing

One or more of the following has
occurred:

The battery is not properly
installed.

A low battery is installed.

A battery fault has occurred.

There is no battery installed
while connected to auxiliary
power.

One or more patient monitoring
parameters have failed self-test
(NIBP, SpO

2

, CO

2

, IBP, or

Temp).

The front panel button self-test
failed.

The speech database self-test
failed.

Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator
continues to flash, remove the unit
from service and contact the
appropriate technical personnel or
the ZOLL Technical Service
Department.

Do Not Use

One or more of the following has
occurred:

The battery is not properly
installed.

No battery is installed and
auxiliary power is not present.

A very low battery (below
software shutdown limit) was
installed.

ECG, defibrillator, or pacer self-
tests have failed, or other critical
self-tests have failed.

Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator
continues to display the Do Not
Use symbol, remove the unit from
service and contact the appropriate
technical personnel or the ZOLL
Technical Service Department.

State

Description

Action