Flashing
One or more of the following has 
occurred:
•
The battery is not properly 
installed.
•
A low battery is installed.
•
A battery fault has occurred.
•
There is no battery installed 
while connected to auxiliary 
power.
•
One or more patient monitoring 
parameters have failed self-test 
(NIBP, SpO
2
, CO
2
, IBP, or
Temp).
•
The front panel button self-test 
failed.
•
The speech database self-test 
failed.
Install a fully charged battery in the 
unit and check the RFU indicator 
again. If the RFU indicator 
continues to flash, remove the unit 
from service and contact the 
appropriate technical personnel or 
the ZOLL Technical Service 
Department.
Do Not Use
One or more of the following has 
occurred:
•
The battery is not properly 
installed.
•
No battery is installed and 
auxiliary power is not present.
•
A very low battery (below 
software shutdown limit) was 
installed.
•
ECG, defibrillator, or pacer self-
tests have failed, or other critical 
self-tests have failed.
Install a fully charged battery in the 
unit and check the RFU indicator 
again. If the RFU indicator 
continues to display the Do Not 
Use symbol, remove the unit from 
service and contact the appropriate 
technical personnel or the ZOLL 
Technical Service Department.
State
Description
Action