Transferring data to a usb device, Clearing the log – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 242
C
HAPTER
21
P
ATIENT
D
ATA
21–4
www.zoll.com
9650-001355-01 Rev. H
Transferring Data to a USB Device
You can transfer a full disclosure log containing patient data from the unit using a USB transfer
device.
Before you begin, insert a USB device into the X Series USB device port.
Figure 21-1.
USB Port
To transfer data through the USB port:
1. Press the power switch to turn the unit on.
2.
Press .
3.
Press .
4.
Press the Transfer Log quick access key (
) (ensure that the USB drive is connected to the
device).
Note:
Do not remove the USB data drive from the X Series unit during transfer.
5.
Use the navigation keys to select Transfer from the Transfer the Log menu.
6.
The green LED on the top of the device turns on while data is transferred to the USB device.
Note:
Wait for the log transfer to complete and for the green light on the top of the X Series
to turn off before removing the USB drive.
Note:
Printing and logging snapshots is disabled during the log transfer.
After transferring data, you must remove the USB drive and reinsert it before attempting
another transfer. If the USB device does not establish communication with the X Series unit, try
powering off the unit and then on again to establish communication.
Warning!
To avoid a possible shock hazard, do NOT make any electrical connections to the USB
port except to connect a USB flash drive while in the vicinity of the patient.
Clearing the Log
You should clear the full disclosure log after transferring data to the USB device or when the
log is nearing its capacity.
Note:
Clearing the log during patient treatment results in the loss of all patient data and
events recorded prior to clearing the log. Clearing the log creates a New Patient record
and all patient-specific parameters (alarm limit, defibrillation energy, etc.) are set to
their default values.
USB Device Port
LOG