Monitoring display options, Capnography (co, Pulse oximetry (spo – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 65
Monitoring Display Options
9650-001355-01 Rev. H
X Series Operator’s Guide
3-3
Capnography (CO
2
)
CO
2
monitoring measures the CO
2
concentration in a patient’s exhaled breath (End Tidal
Carbon Dioxide --EtCO
2
). CO
2
monitoring can also measure a patient’s breath rate and the
CO
2
concentration in the gasses supplied to intubated patients (Fractional Inspired Carbon
Dioxide -- FiCO
2
). Since FiCO
2
represents the amount of CO
2
present during inhalation, it also
serves as an indicator for rebreathing in non-intubated patients. CO
2
monitoring can be used for
both intubated and non-intubated patients.
The EtCO
2
, breath rate, and FiCO
2
measurements appear in a labeled (
EtCO2
) numeric
display. The EtCO
2
and FiCO
2
measurements can appear as values given in millimeters of
mercury (mmHg). You can also specify that the X Series unit display a CO
2
capnogram in the
waveform trace display area.
Pulse Oximetry (SpO
2
)
Pulse Oximetry monitoring measures the oxygen saturation (SpO
2
) of arterial blood at a
peripheral site such as a finger or toe. If optional features SpCO and SpMet are installed, it also
monitors carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet).
SpO
2
monitoring determines the ratio of oxygenated hemoglobin to total hemoglobin in arterial
blood and displays this ratio as percent SpO
2
in a labeled (
SpO2
) numeric display. If optional
features SpCO and SpMet are installed, these values alternate and display as
SpCO
and
SpMet
under the SpO
2
display. You can also specify that the X Series unit display an SpO
2
plethysmograph in the waveform trace display area.
Monitoring Display Options
The X Series unit gives you great flexibility in how you display a patient’s vital signs
information. By pressing the Display/Home button (
) on the front panel, you can
successively display the patient’s vital signs information in these three windows:
•
Waveform Display window, which initially displays an ECG waveform trace and numeric
displays for each monitoring function.
•
Trends Status window, which displays a report listing vital signs measurements that the X
Series unit logs automatically, and the primary ECG waveform trace.
•
Large Numerics Display window, on which large numeric displays of all vital signs
measurements appear.
The Waveform Display window appears when you power on the X Series unit. Initially, the
Waveform Display window displays a single ECG waveform trace. All other monitored values
appear in numeric display areas at the bottom of the screen: