beautypg.com

Ensuring correct cuff inflation settings – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 114

background image

C

HAPTER

8 M

ONITORING

N

ON

-I

NVASIVE

B

LOOD

P

RESSURE

(NIBP)

8-8

www.zoll.com

9650-001355-01 Rev. H

Caution

Using a cuff that is loosely applied or too small results in measurements higher than the
patient’s actual blood pressure.

Using a cuff that is too large results in values lower than the patient’s actual blood pressure.

Ideally, the cuff should be at the same level as the heart. Cuff placement substantially above
or below heart level will result in blood pressure measurements that are erroneously low or
high.

The following illustrates one possible cuff placement for adult/pediatric patients and, to the
right, possible cuff placements for neonates:

Figure 8-3 Applying Cuff to the Patient

Ensuring Correct Cuff Inflation Settings

Before taking an NIBP measurement for a new patient, ensure that the cuff inflation settings are
appropriate for that patient.

Check that the correct patient mode is selected. The initial Cuff Inflation Target (CIT) pressure
is dependent on the patient type and the configured CIT preset. The default and configurable
CIT presets for each patient type are indicated in the following table (default values are in
bold).

Note:

You can configure the CIT pressure from the Setup>NIBP menu.

Adult

Pediatric

Neonate

120 mmHg

80 mmHg

60 mmHg

140 mmHg

90 mmHg

70 mmHg

160 mmHg

100 mmHg

80 mmHg

180 mmHg

110 mmHg

90 mmHg

200 mmHg

120 mmHg

100 mmHg

220 mmHg

130 mmHg

110 mmHg