Zeroing the transducer – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 151
Zeroing the Transducer
9650-001355-01 Rev. H
X Series Operator’s Guide
11-3
Figure 11-1 Plugging the Transducer into the X Series
When you plug the transducer cable into the unit, the message ZERO PROBE appears in the
numeric display window for that IBP channel.
Zeroing the Transducer
To ensure that the X Series unit measures pressure accurately, you must zero the transducer
before each use. If you change or disconnect a transducer, you must zero the new transducer
before use. If you move a transducer to a different monitor, you must zero the transducer again,
even if you have already done so on another unit. In addition to the procedure below, follow the
transducer manufacturer’s Directions for Use and your organization’s established clinical
protocol.
Follow these steps when zeroing a transducer:
1. Place the transducer at the same height as the patient’s left atrium.
2.
Close the transducer stopcock to the patient.
3.
Open the transducer’s venting stopcock to atmospheric air.
4.
Allow the transducer a few seconds to settle.
5.
Use the navigation keys to highlight and select the tranducer’s IBP channel and display the
IBP channel’s parameter control panel: