Using the spo, Parameter control panel, Selecting the spco and spmet monitoring – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 144
C
HAPTER
10 P
ULSE
CO-O
XIMETRY
(S
P
O2, SPCO,
AND
S
P
M
ET
)
10-10
www.zoll.com
9650-001355-01 Rev. H
Setting Upper and Lower SpCO and SpMet Alarm Limits
The following table lists the default SpCO and SpMet limits for adult, pediatric, and neonate
patients, and gives the range in which you can set these limits.
Using the SpO
2
Parameter Control Panel
Use the navigation keys to highlight and select the SpO
2
numeric display to display the
SpO2 Parameter Control Panel, on which you can select options that optimize the use of SpO
2
for the patient:
Figure 10-4 SpO2 Parameter Control Menu
Selecting the SpCO and SpMet Monitoring
If SpCO and SpMet are installed, the X Series unit allows you enable or disable SpCO and
SpMet monitoring.
Patient Type
SpCO Limit
Default
SpCO Limit Range
SpMet Limit
Default
SpMet Limit Range
Adult
Lower: 0%
Upper: 10%
Lower: 0- 98%
Upper: 2 - 100%
Lower: 0%
Upper: 3%
Lower: 0- 98%
Upper: 2 - 100%
Pediatric
Lower: 0%
Upper: 10%
Lower: 0 - 98%
Upper: 2 - 100%
Lower: 0%
Upper: 3%
Lower: 0- 98%
Upper: 2 - 100%
Neonate
Lower: 0%
Upper: 10%
Lower: 0 - 98%
Upper: 2 - 100%
Lower: 0%
Upper: 3%
Lower: 0- 98%
Upper: 2 - 100%
SpO2
Lower
Upper
SpO2 Alarm
85
100
SpCO Alarm
SpMet Alarm
On
On
8 seconds
Normal
Off
SpCO Monitoring
SpMet Monitoring
Averaging Time
Sensitivity
HR/PR Tone
0.0
0
10
3.0