beautypg.com

Chapter 8, Monitoring non-invasive blood pressure (nibp) – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 107

background image

9650-001355-01 Rev. H

X Series Operator’s Guide

8-1

Chapter 8

Monitoring Non-Invasive Blood Pressure (NIBP)

The X Series NIBP input is Type CF defibrillator proof.

This chapter describes how to use the X Series unit to perform Non-Invasive Blood Pressure
(NIBP) measurements using an inflatable cuff to measure arterial pressure.

The X Series uses the enhanced NIBP features of Welch Allyn’s SureBP and patented Smartcuf
motion-tolerant technology.

The X Series unit must perform ECG monitoring when using Smartcuf.

The SureBP monitoring software allows the X Series unit to take an NIBP measurement as the
cuff is inflating, which saves time -- the measurement takes about 15 seconds -- and improves
patient comfort.

The Smartcuf monitoring software enables the X Series unit to make accurate NIBP
measurements in the presence of extreme artifact, weak pulses, and some dysrhythmias by
synchronizing the NIBP measurements with the patient’s R-wave.

Warning!

Observe the patient’s limb periodically to ensure that circulation is not impaired for
a prolonged period of time.

Never use the X Series to monitor NIBP on one patient while simultaneously
monitoring ECG on another patient.

If a non-invasive blood pressure measurement is suspect, repeat the measurement. If
you are still uncertain about the measurement, use another method.