Setting co, And respiration rate alarms, Enabling/disabling alarms and setting co – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 130: Alarm limits
C
HAPTER
9 M
ONITORING
CO2
9-8
www.zoll.com
9650-001355-01 Rev. H
Setting CO
2
and Respiration Rate Alarms
The X Series unit sounds alarms whenever measurements are outside set limits for the
following:
•
High and Low EtCO
2
•
High and Low Respiration Rate (in Breaths/Minute)
•
High FiCO
2
•
No Breath Alarm
Note:
Use the Resp Parameter Control Panel to set the duration of the No Breath Alarm, or to
disable it (see “Using the Resp Parameter Control Panel” on page 7-4).
Enabling/Disabling Alarms and Setting CO
2
Alarm Limits
You can enable (or disable) CO
2
alarms and set upper and lower alarm limits through the
Alarms quick access key (
) or the CO2 Parameter Control Panel.
To configure CO
2
alarms through the Alarms quick access key:
1. Press the More quick access key (
).
2.
Press .
3.
Press the Limits quick access key (
).
4.
Use the navigation buttons to select appropriate alarm menu. For CO
2
, the alarm menus are:
EtCO2 Alarm, FiCO2 Alarm, or RR/BR Alarm.
5.
On the alarm settings menu, use the navigation keys to select the fields that you want to
change. The fields are
•
Upper Limit Enable
•
Lower Limit Enable (only EtCO2 and RR/BR Alarms)
•
Upper Limit
•
Lower Limit (only EtCO2 and RR/BR Alarms)
6.
When you are finished changing values on the alarm menu, select the Backarrow key to exit
the menu.
Limits