General – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 304
Specifications
A-18
www.zoll.com
9650-001355-01 Rev. H
General
Weight:
10.6 lbs.without battery and paper
11.7 lbs.with battery and paper
Dimensions:
Without Handle: 8.9” x 8.75” x 7.9”
With Handle: 8.9” x 10.4” x 7.9”
Operating:
Temperature: 0 to 50° C
Humidity: 15 to 95% RH (non-condensing)
Vibration:
•
EN ISO 9919 (per IEC 60068-2-64)
•
RTCA/DO-160G (multiple helicopter frequencies)
•
EN 1789 for ambulance
Shock: IEC 60068-2-27, 100g, 6ms half sine
Bump: IEC 60068-2-29
Drop: EN 1789, 30” functional drop
IEC 60601-1, Tested at 2 meters
Altitude: -170 M to 4572 M (-557 feet to 15,000 feet)
Transport and Storage:
Temperature: -30 to 70°C
Humidity: 15 to 95% RH (non-condensing)
Atmospheric pressure: 572 mbar to 1034 mbar
Shock/vibration: MIL STD 810G, Method 514.6, 4.4.2, Procedure II
Safety Classification: Class 1 and internal power per IEC/EN 60601-1
Enclosure Protection:
Solid Foreign Object: IEC 60529, IP5X
Water: IEC 60529, IPX5
AC Auxiliary Operating Power:
8300-0004
Input:
100-240V
50-60 Hz, 2A
100-115V
400 Hz, 2A
Output: 14.5V
4.15A
80W (peak)
IP Rating: IP23
Note:
The X Series device may not perform to specifications when stored at the upper or
lower extreme limits of storage temperature and immediately put into use.