beautypg.com

General – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 304

background image

Specifications

A-18

www.zoll.com

9650-001355-01 Rev. H

General

Weight:

10.6 lbs.without battery and paper

11.7 lbs.with battery and paper

Dimensions:

Without Handle: 8.9” x 8.75” x 7.9”

With Handle: 8.9” x 10.4” x 7.9”

Operating:

Temperature: 0 to 50° C
Humidity: 15 to 95% RH (non-condensing)

Vibration:

EN ISO 9919 (per IEC 60068-2-64)

RTCA/DO-160G (multiple helicopter frequencies)

EN 1789 for ambulance

Shock: IEC 60068-2-27, 100g, 6ms half sine

Bump: IEC 60068-2-29

Drop: EN 1789, 30” functional drop

IEC 60601-1, Tested at 2 meters

Altitude: -170 M to 4572 M (-557 feet to 15,000 feet)

Transport and Storage:

Temperature: -30 to 70°C

Humidity: 15 to 95% RH (non-condensing)

Atmospheric pressure: 572 mbar to 1034 mbar

Shock/vibration: MIL STD 810G, Method 514.6, 4.4.2, Procedure II

Safety Classification: Class 1 and internal power per IEC/EN 60601-1

Enclosure Protection:

Solid Foreign Object: IEC 60529, IP5X

Water: IEC 60529, IPX5

AC Auxiliary Operating Power:

8300-0004

Input:

100-240V

50-60 Hz, 2A

100-115V

400 Hz, 2A

Output: 14.5V

4.15A

80W (peak)

IP Rating: IP23

Note:

The X Series device may not perform to specifications when stored at the upper or
lower extreme limits of storage temperature and immediately put into use.