Chapter 14, Lead ecg interpretive analysis – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 175
9650-001355-01 Rev. H
X Series Operator’s Guide
14-1
Chapter 14
12-Lead ECG Interpretive Analysis
The X Series 12-Lead input is Type CF defibrillator proof.
This chapter describes how to use the X Series unit to monitor 12-Lead ECG for adult and
pediatric patients, and how to display 12-Lead ECG Interpretive Analysis information for adult
patients.
X Series 12-Lead ECG Monitoring provides simultaneous acquisition and storage of
12-Lead information for adult and pediatric patients, and, when enabled, post-acquisition
Interpretive Analysis for adult patients.
Warning!
•
12-Lead ECG monitoring is intended for the recording of 12-lead ECG signals from
adult and pediatric patients in the supine, resting position -- always ensure that the
patient is kept motionless during 12-lead ECG signal acquisition and analysis. Use of
the device to acquire ECG signals from moving or shaking patients may produce noisy
signals that are difficult to interpret.
•
The 12-Lead Interpretive algorithm's interpretive statements are designed to enhance
the diagnostic process. They are no substitute for the qualified judgment of a properly
trained clinician. As with any diagnostic test, always give consideration to patient
symptoms, history, and other relevant factors.
•
12-Lead Interpretive Analysis is for use with only adult patients.
•
Excessive body hair or wet, sweaty skin may interfere with electrode adhesion. Remove
the hair and/or moisture from the area where the electrode is to be attached.
•
Remove ECG electrodes from their sealed package immediately prior to use. Using
previously opened or out-of- date electrodes may degrade the ECG signal quality.
•
Monitoring electrodes may become polarized during defibrillator discharge, causing
the ECG waveform to briefly go off screen. ZOLL Medical Corporation recommends