Configuring NIBP Alarms and Settings
9650-001355-01 Rev. H
X Series Operator’s Guide
8-13
By default, Smartcuf is Enabled.
There are situations where you may choose to disable Smartcuf. These may include
•
Very extreme motion artifact
•
Certain types of arrhythmias
•
Situations that prevent getting a ECG signal.
NIBP measurements can still be performed with Smartcuf disabled.
Start/Stop TurboCuf
Selecting Start TurboCuf starts Short-term Automatic (STAT) NIBP measurements. The X
Series unit begins its first NIBP measurement, after which it continues to perform as many
NIBP measurements as possible over a 5-minute period.
Select Stop TurboCuf
to immediately stop STAT measurements.
WARNING!
Repeated use of STAT measurements on the same patient over a short time interval can
affect blood pressure readings, limit circulation to the limb, and cause injury to the
patient.
Setting the NIBP Display Format
You can specify the appearance of the NIBP numeric display format. The display format allows
you to choose whether or not you will display MAP measurement (
M
) with the
Systolic (
S
) and Diastolic (
D
) measurements in one of the following formats:
•
S/D
•
S/D (M)
(default display)
•
(M) S/D
Note:
If a MAP alarm is active when the
S/D
format is selected, the display format will be
S/D (M)
until you clear the MAP alarm.