Changing the trends status window display, Continuous waveform recording – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 75
Changing the Trends Status Window Display
9650-001355-01 Rev. H
X Series Operator’s Guide
4-3
Changing the Trends Status Window Display
By default, the Trends status window displays all logged trend information. It displays the
numeric information for all monitoring functions, which the unit logs at a user-configurable
interval, and when you take NIBP measurements, when a patient alarm occurs, and when you
press .
To configure the display of the Trends status window, press (
) and press the Trend Settings
quick access key (
) to display the Trends Settings control panel. On the Trends Settings
control panel, select
Trend Display Format
to specify the following monitored vital signs that
appear in the Trends status window:
Continuous Waveform Recording
Continuous waveform recording allows you to record continuous waveforms to a full
disclosure case log for the patient being monitored. This feature needs to be enabled by a
supervisor. The supervisor can access this feature by pressing the Setup quick access key (
),
and then selecting Supervisor>Log>Waveform Recording. In this menu, the following settings
can be customized:
•
Record Displayed Waveforms – Record only the top waveform that is displayed or record
all waveforms that are displayed.
•
Record Additional Waveforms – Always record the CO
2
waveform or always record the
IBP and the CO
2
waveforms.
There is an additional waveform that can be enabled in the Waveform Recording menu. This
option, Record Pads Imp. Waveform, measures the patient’s impedance between
defibrillation electrodes.
Note:
When Continuous waveform recording is disabled, no waveforms (except snapshots)
are recorded.
The full disclosure case can, at a minimum, concurrently store the following information:
•
32 monitor snapshots
•
500 non-ECG events
•
24 hours of continuous ECG (4 waveforms), Capnography, IBP (3 channels), and Pads
Impedance
The actual information that is stored depends on usage. Also, the specific combination of stored
continuous waveform data depends on how the waveform recording settings are configured in
the Supervisor menu.
Trend Format
Vital Signs Displayed
Resp
HR, SpO2, RR, EtCO2, FiCO2
NIBP
HR, SpO2, NIBP, RR
IBP1
HR, SpO2, IBP1, RR
IBP2
HR, SpO2, IBP2 RR
IBP3
HR, SpO2, IBP3, RR
Temp
HR, SpO2, T1, T2,
T
LOG