ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 320
Specifications
A-34
www.zoll.com
9650-001355-01 Rev. H
Recommended Separation Distances from RF Equipment for the X Series
IBP Non Life-Supporting Function
Non life-supporting functions of the X Series include
:
12-lead ECG, EtCO
2
, SpO
2,
SPCO,
SpMet, NIBP, IBP, Temp, Impedance Respiration.
Recommended separation distances between portable and mobile RF
communications equipment and the X Series
The non life-supporting functions of the X Series are intended for use in the electromagnetic environment in
which radiated RF disturbances are controlled. The customer or user of the X Series can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the X Series as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output power of
equipment
(in watts)
Separation distance according to frequency of transmitter
(in meters)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in
meters can be determined using the equation applicable to the frequency of the transmitter, where
P
is the
maximum output power rating of the transmitter in watts according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
d =1.2
P
d =1.2
P
d =2.3
P