ZOLL AutoPulse Rev D Quick Reference Guide User Manual
For adult, Non-traumatic cardiac arrest, Troubleshooting

1. Power up AutoPulse
2. Close chest bands
3. Press CONTINUE (green button)
4. Press START (green button) to
begin compressions
To pause or stop operation
press STOP (orange button)
• Remove ALL clothing from torso
(both front and back) to ensure
skin-to-platform contact
• Align armpits onto yellow line
on Platform
Do not twist the
chest bands
Maintain chest bands at
90 degrees to Platform
and free of obstructions
For Fault/User Advisory
• Lift up and fully extend both chest bands.
• Check both lateral and vertical patient alignment.
• Verify that chest bands are not twisted,
are 90 degrees to the Platform and are
free of obstructions.
• Press RESTART (green button) and follow
on-screen instructions to begin compressions.
If you cannot rectify problem immediately open chest
bands and revert to manual
CPR
.
Troubleshooting
Follow all local protocols
and procedures.
Review User Guides and
complete in-service training.
LifeBand
®
instructions on other side
P/N 10596-001 Rev. 10
AutoPulse
®
Quick Reference Guide
For Adult
(>–18 years)
Non-Traumatic Cardiac Arrest
Maximum Patient Weight 300 lbs.