Defibrillator/Pacing Test with Hands-Free Therapy Electrodes
9650-001355-01 Rev. H
X Series Operator’s Guide
24-3
Defibrillator/Pacing Test with Hands-Free Therapy
Electrodes
Note:
If a low battery message appears during any of this testing, the battery is close to
depletion and should be replaced or recharged.
Warning!
Keep hands, fingers, and other conductive materials away from paddle electrode plates
when discharging the defibrillator, or pacing.
Function
Response
1
Connect the external AC adapter to a working
AC outlet and to the X Series rear panel.
Verify that the green Auxiliary Power LED
illuminates on the X Series front panel.
2
Insert a battery into the unit (if a battery is not
already in the slot).
Verify that the Battery Charge LED on the X
Series front panel illuminates yellow. When the
battery pack has been fully charged, the
battery status indicator light is green.
3
Press the unit's power on switch.
Verify the unit issues two audio beeps and that
the red, yellow, and green Visual Alarm
indicators illuminate for 2 to 3 seconds.
Note: The yellow LED may remain lit after the
unit powers on.
The X Series unit performs a number of power
on self tests.
Verify that the unit displays SELF TEST
PASSED.
If the message SELF TEST FAILED appears
on the display, contact the appropriate
technical personnel or the ZOLL Technical
Service Department.
4
Disconnect the external AC adapter from the
X Series rear panel.
Verify that the unit continues to operate without
interruption on battery power and that the
displayed battery icon indicates that a fully
charged battery is installed.
Verify that no error messages are displayed.
5
With no ECG cables attached to the unit,
select Lead I, II, or III as the waveform display
source.
Verify that a LEAD FAULT message displays in
the waveform window.
6
Select Pads as the waveform display source
and connect the multifunction cable to the
unit. (Do not connect the test connector to the
MFC.)
Verify that a CHECK THERAPY
ELECTRODES message displays in the
waveform window.
7
Connect the multi-function cable (MFC) to the
test connector.
Verify that the CHECK THERAPY
ELECTRODES message is replaced by a
SHORT DETECTED message.