Applying the cuff to the patient – ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 113

background image

Applying the Cuff to the Patient

9650-001355-01 Rev. H

X Series Operator’s Guide

8-7

2.

Insert the plastic connectors on the NIBP hose into the cuff hose connector, and twist the
connectors until they lock. The connectors are the same; you can insert either of the plastic
connectors into either of the cuff hose connectors in any order.

Applying the Cuff to the Patient

To apply the cuff to the patient:

1. Ensure the patient is sitting or lying down with the limb relaxed, extended, and placed on a

smooth surface for support.

2.

Squeeze as much air from the cuff as possible before placing it on the patient.

3.

Place the cuff 2 to 5 cm above the elbow crease or 5 to 10 cm (1.9 to 3.9 in.) above the knee
crease.

Warning!

Do not place the NIBP cuff on the same arm or leg as an SpO

2

sensor. Inflation of the

cuff causes the SpO

2

monitor to read incorrectly. Also, do not attach the cuff to a limb

being used for IV infusion. Cuff inflation might block the infusion, causing harm to the
patient.

4.

Adjust the cuff so that the artery marker on the cuff is over the artery, pointing to the hand or
foot.

5.

Check that the cuff ends between the range lines marked on the cuff.

6.

If they do not line up, use a different size cuff.

7.

Wrap the deflated cuff snugly around the limb without impeding blood flow.

8.

Ensure that the hose is routed to avoid kinking or compression.