Defibrillator testing with external paddles – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 279

Defibrillator Testing with External Paddles
9650-001355-01 Rev. H
X Series Operator’s Guide
24-5
Defibrillator Testing with External Paddles
Prior to testing external defibrillator paddles with the X Series unit, complete the testing 
described in “Defibrillator/Pacing Test with Hands-Free Therapy Electrodes” on page 24-3. 
Note:
If a low battery message appears during any of this testing, the battery is close to 
depletion and should be replaced or recharged.
Warning!
Keep hands, fingers, and other conductive materials away from paddle electrode plates 
when discharging the defibrillator, or pacing.
Function
Response
1
Press the X Series power switch to turn the 
unit on.
Verify that the unit displays the SELF TEST 
PASSED message.
2
Connect the MFC to the defibrillator and 
disconnect the patient end of the MFC from 
any attached accessory (paddles or test 
connector). Select Pads as the waveform 
display source.
Verify that a CHECK THERAPY 
ELECTRODES message displays in the 
waveform window
3
Connect the paddles set to the X Series MFC 
cable. Do not place the paddles electrodes in 
contact with each other.
Verify that the CHECK THERAPY 
ELECTRODES message changes to APPLY 
PADDLES TO PATIENT.
4
Press the RECORDER ON/OFF button on 
the Sternum paddle.
Verify that the recorder begins printing.
5
Press the RECORDER ON/OFF button again. Verify that the recorder stops printing.
6
Press the Apex and Sternum paddle 
electrodes together (face to face) maximizing 
the contact area between the electrodes. 
Verify that the APPLY PADDLES TO PATIENT 
message changes to SHORT DETECTED.
7
Press the ENERGY SELECT (+) button on 
the Sternum paddle.
Verify that the defib window displays and the 
following message appears:
PADDLE SHORT DETECTED
SELECT 30 J TO TEST
8
Press the ENERGY SELECT (+) button 
again.
Verify that the selected energy increases to the 
next highest level.
9
Repeatedly press the ENERGY SELECT (-) 
button on the Sternum paddle until 30 Joules 
is selected.
Verify that the defibrillator window shows
30 J as the selected energy.
10 Press the CHARGE button on the Apex
paddle. (Press the CHARGE button twice if 
the defibrillator window has disappeared.)
Verify that a periodic tone sounds while the 
defibrillator is charging. At the completion of 
the charge cycle, the tone is continuous. The 
charge time should be less than 7 seconds.
Verify that the CHARGE light on the Apex 
handle illuminates when the defibrillator is 
charged and ready to deliver energy.
11
While the defibrillator is charged, press the 
paddles electrodes together and then press 
and hold the SHOCK button on the Apex 
paddle (only).
Verify that the defibrillator does not 
discharge.