Alarms
Heart Rate Alarms:
Audible: 5 pulse, 900 Hz tone, with a PW of 125 msec, a PRI of 250 msec, and a repetition
interval of 15 seconds.
Visual: Heart Rate Alarm causes the heart rate to be displayed in red, with a white background.
The red device status LED will flash a rate of 1.7 Hz.
Lead Fault Alarm:
Audible: 3 pulse, 500 Hz, triplet tone with a PW of 200 msec, a PRI of 200 msec. The lead
fault tone repeats at a repetition interval of 30 seconds.
Visual: Lead Fault condition causes a LEAD FAULT message to be displayed on the trace
along with a dashed line the width of the trace.
Physiological Alarms (NIBP, SpO
2
, Resp, CO
2
, IP & Temp):
Audible: Same as Heart Rate Alarm
Visual: Physiological alarms cause the alarming parameter to be displayed in Red with a white
background. The red device status LED will flash at a rate of 1.7 Hz.
Silence Duration: 90 seconds.
Invalid Operation Alert Tone:
A short, low-pitched tone is audible when a selected control button is unavailable for use or an
invalid entry is detected. Tone frequency is 160 Hz. Duration is 250 msec.
Delays (hold off time):
Heart rate/pulse rate:
•
i
f source is ECG, 5 seconds
•
if source is SpO
2
, 10 seconds
•
if source is IBP, 5 seconds
•
if source is NIBP, no hold off
SpO
2
, SpCO, and SpMet Saturation: 10 seconds
EtCO
2
: 7 seconds
FiCO
2
: 5 seconds
IBP (Systolic, Diastolic, Mean): 2 seconds
Temperature: 2 seconds