Warnings -- SpO
2
, Oximeter Sensor
•
Use only ZOLL-approved Masimo oximeter sensors. Other manufacturers’ sensors
might not perform properly with the X Series oximeter.
•
Tissue damage can result from incorrect application or use of a sensor (for example,
wrapping the sensor too tightly, applying supplemental tape, failing to periodically
inspect the sensor site). To ensure skin integrity, correct positioning, and sensor
adhesion, inspect the sensor site as directed in the Directions for Use provided with the
sensor.
•
Do not use damaged sensors or cables.
•
Do not use a sensor with exposed optical components.
•
Do not sterilize a sensor by irradiation, steam, or ethylene oxide. Refer to the cleaning
instructions in the Directions for Use for reusable Masimo sensors.
•
Do not allow the sensor to remain on the same site for a prolonged period, especially
when monitoring neonates. Check the application site at regular intervals (at least
every 2 hours) and change the site if any compromise in skin quality occurs.
•
Do not attach the oximeter sensor to a limb being monitored with a pressure cuff or
with restricted blood flow.
•
A poorly applied sensor might give incorrect saturation readings. A weak pulse signal
on the display might indicate a poorly applied sensor or a poorly chosen monitoring
site.
•
Choose a site with sufficient perfusion to ensure accurate oximetry values.
•
Certain nail aberrations, nail polish, fungus, and so on might cause inaccurate
oximetry readings. Remove any nail polish or move the sensor to an unaffected digit.
•
Exposure to high ambient light from surgical lamps (especially those with a xenon light
source), bilirubin lamps, fluorescent lights, infrared heating lamps, or direct sunlight
can affect the accuracy of oximetry readings. To prevent interference from ambient
light, ensure that the sensor is properly applied. If necessary, cover the sensor with
opaque material.
•
Always remove the sensor from the patient and completely disconnect the patient from
the pulse CO-Oximeter before bathing the patient.
•
Do not attempt to reprocess, recondition or recycle any Masimo sensors or patient
cables as these processes may damage the electrical components, potentially leading to
harm.Do not modify or alter the sensor in any way. Alterations or modification may
affect performance and/or accuracy.
•
To avoid cross contamination only use Masimo single use sensors on the same patient.