ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 154

background image

C

HAPTER

11 M

ONITORING

I

NVASIVE

P

RESSURES

(IBP)

11-6

www.zoll.com

9650-001355-01 Rev. H

Enabling/Disabling IBP Alarms and Setting Alarm Limits

When enabled, the X Series unit sounds alarms whenever IBP measurements are outside set
limits for the following:

High and Low Systolic Pressure

High and Low Diastolic Pressure

High and Low Mean Arterial Pressure (MEAN)

To enable (or disable) IBP alarms and set Upper and Lower alarm limits and IBP channel, you
can either use the navigation buttons to highlight and select the IBP channel display or:

1. Press the More quick access key (

).

2.

Press .

3.

Press the Limits quick access key (

).

4.

Use the navigation buttons to highlight and select the appropriate alarm menu selection. For
each IBP channel, there are alarm menus for systolic (Systolic Alarm), diastolic (Diastolic
Alarm) and mean arterial pressure (Mean Alarm) alarm settings. On the selected Alarm
Settings menu, use the navigation keys to select the fields that you want to change. The fields
are

Upper Limit Enable

Lower Limit Enable

Upper Limit

Lower Limit

5.

When you are finished changing values on the alarm menu, navigate to the Backarrow key
to exit the menu.

Setting Upper and Lower Systolic (SYS) Alarm Limits

Initially, the IBP Systolic Alarm Settings menu specifies that the IBP systolic pressure alarms
are enabled (

ON

) or disabled (

OFF

)

,

and displays the default upper and lower systolic limits.

The upper and lower limits can be

ON

or

OFF

(default is

OFF

). The following table lists the

default IBP Systolic limits for adult, pediatric, and neonate patients, and gives the range in
which you can set these limits:

Limits

Patient Type

IBP Systolic Limit Default

IBP Systolic Limit Range

Adult

Lower: 75 mmHg

Upper: 220 mmHg

Lower: -30 to 298 mmHg

Upper: -28 to 300 mmHg

Pediatric

Lower: 75 mmHg

Upper: 145 mmHg

Lower: -30 to 298 mmHg

Upper: -28 to 300 mmHg

Neonate

Lower: 50 mmHg

Upper: 100 mmHg

Lower: -30 to 298 mmHg

Upper: -28 to 300 mmHg