ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 289
Defibrillator
9650-001355-01 Rev. H
X Series Operator’s Guide
A-3
* For all energy levels, accuracy is equal to either ±15% or 3 joules, whichever is greater.
The X Series Rectilinear Biphasic waveform employs the same first and second phase timing,
similar first and second phase currents/voltages, and essentially the same mechanisms for
controlling defibrillation waveshape as the ZOLL R Series
®
defibrillator. The R Series and X
Series defibrillation waveforms are considered substantially equivalent.
Figures A-1 through A-20 show the Rectilinear Biphasic waveforms that are produced when
the X Series defibrillator is discharged into loads of 25, 50, 75, 100, 125, 150, and 175 ohms at
each energy setting.
Table A-2.
Delivered Energy at Every Defibrillator Setting into a Range of Loads
Selected
Energy
Load
25
50
75
100
125
150
175
Accuracy*
1 J
1 J
1 J
1 J
1 J
1 J
1 J
1 J
±15%
2 J
1 J
2 J
2 J
2 J
2 J
2 J
2 J
±15%
3 J
2 J
3 J
3 J
3 J
3 J
3 J
3 J
±15%
4 J
3J
4 J
4 J
5 J
5 J
5 J
4 J
±15%
5 J
3 J
5 J
6 J
6 J
6 J
6 J
6 J
±15%
6 J
4 J
6 J
7 J
7 J
7 J
7 J
7 J
±15%
7 J
5 J
7 J
8 J
8 J
8 J
8 J
8 J
±15%
8 J
5 J
8 J
9 J
9 J
10 J
9 J
9 J
±15%
9 J
6 J
9 J
10 J
11 J
11 J
11 J
10 J
±15%
10 J
7 J
10 J
12 J
12 J
12 J
12 J
12 J
±15%
15 J
10 J
16 J
17 J
18 J
18 J
18 J
17 J
±15%
20 J
14 J
21 J
23 J
24 J
24 J
24 J
23 J
±15%
30 J
21 J
32 J
35 J
36 J
37 J
36 J
35 J
±15%
50 J
35 J
54 J
59 J
61 J
62 J
61 J
59 J
±15%
70 J
49 J
76 J
83 J
85 J
87 J
86 J
83 J
±15%
85 J
60 J
92 J
101 J
104 J
106 J
104 J
101 J
±15%
100 J
71 J
109 J
119 J
122 J
125 J
123 J
119 J
±15%
120 J
85 J
131 J
143 J
147 J
150 J
147 J
143 J
±15%
150 J
107 J
164 J
180 J
183 J
188 J
184 J
179 J
±15%
200 J
142 J
230 J
249 J
253 J
269 J
261 J
260 J
±15%