Switching to Manual Mode Operation

Press the Manual Mode quick access key on the front panel of the unit to enter the Manual
mode of operation.

Using the navigation keys, select the four digits in the Manual Mode pass code. Press SAVE
when you are finished. Once you have entered your pass code, you will be able to enter Manual
mode.

Note:

If the unit has not been configured to enter a pass code, the message Exit to Manual
Mode
is displayed. Use the navigation keys to select Yes to enter the Manual mode of
operation. If you do not press Yes within 10 seconds, the unit will revert back to AED
operation.

When changing from AED mode to Manual mode, the current selected energy level is
maintained.

Note:

To transfer back to AED mode from Manual mode, power down the unit for more than
30 seconds and less than two minutes, then power it back up. If you wait longer than
two minutes, the unit will reset the settings to the defaults and treat the case as a new
patient.