9650-001355-01 Rev. H

X Series Operator’s Guide

5-1

Chapter 5

Alarms

The X Series unit supports the detection and indication of patient alarms and technical alerts.

A patient alarm is any alarm condition that is caused by a monitored patient-related variable,
such as a measured vital sign that falls outside of a configured alarm limit. You can configure
patient alarm limits for each of the physiologic monitoring functions.

A technical alert is monitored equipment-related variable that the X Series unit can detect, such
as a disconnected sensor, internal diagnostics failures, and so on. Technical alerts are always
enabled and are not user-configurable
.

Patient alarms are always classified as high-priority alarms. Technical alerts are classified as
medium or low priority alarms.

Alarm conditions from patient alarms and technical alerts are stored in the Event Log and
retained with normal power down or total loss of power.

Visual Alarm Indicators

In addition to status messages that appear on the display, the X Series unit lights the red or
yellow LED on the front panel to indicate the priority level of the highest-priority active alarm.
The X Series LEDs indicate the priority level of the highest-priority active alarm as shown in
the following table.

Active Alarm Priority

Visual Alarm Indicator

High Priority -- Patient Alarm

Flashing Red LED

Medium Priority -- Technical Alert

Flashing Yellow LED

Low Priority -- Technical Alert

Continuous Yellow LED