Cpr dashboard, Constant metronome disabled – ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 229
CPR Dashboard
9650-001355-01 Rev. H
X Series Operator’s Guide
19–3
Constant Metronome Disabled
AED Mode
•
If Enable Metronome is set to Always or AED Mode Only, the metronome is active when
chest compressions are first detected during the CPR interval.
•
If set to Never, the metronome is silent.
Advisory/CPR Protocol or Manual Mode
•
If Enable Metronome is set to Always, the metronome is active only when the reported
compression rate falls below 80 cpm during the CPR period. If the chest compression sensor
is connected, the metronome is suspended after compressions have not been detected for at
least 2 seconds.
•
If set to Never or AED Mode Only, the metronome is silent.
The following table indicates when the CPR metronome is active.
CPR Dashboard
Whenever CPR-D-padz or OneStep electrodes
are connected to the X Series defibrillator and it
senses compression, the unit illuminates the CPR Dashboard in the lower center portion of the
display. This field includes the CPR feedback indicators described in the following sections.
Enable
Metronome
Setting
AED Mode
Advisory/CPR
Protocol Mode
Manual Mode
Constant Metronome Enabled
AED Mode
Only
On when CPR detected
Off when CPR Idle
Off
Off
Always
On when CPR detected
Off when CPR Idle
On when CPR detected
Off when CPR Idle
On when CPR detected
Off when CPR Idle
Never
Off
Off
Off
Constant Metronome Disabled
AED Mode
Only
On at start of CPR period
Off
Off
Always
On at start of CPR period
On only at <80
compressions per minute
On only at <80
compressions per minute
Never
Off
Off
Off
CPR Release Indicator
Chest Compression
Indicator