ZOLL X Series Monitor Defibrillator Rev H User Manual
Page 315

Electromagnetic Compatibility Guidance and Manufacturer’s Declaration
9650-001355-01 Rev. H
X Series Operator’s Guide
A-29
Electromagnetic Compatibility Guidance and 
Manufacturer’s Declaration
In-Flight Use (RTCA/DO-160):
The X Series unit complies with RTCA/DO-160, Environmental Conditions and Test 
Procedures for Airborne Equipment, using the methods in Section 21, Category M for Radiated 
and Conducted Radio Frequency Energy.
Guidance and manufacturer’s declaration – electromagnetic emissions
The X Series unit is intended for use in the electromagnetic environment specified below. The customer or the 
user of the X Series unit should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF Emissions CISPR 11
Group 1
The X Series unit uses RF energy only for its internal 
function. Therefore, its RF emissions are very low and are 
not likely to cause any interference in nearby electronic 
equipment. 
RF emissions CSPR 11
Class B
Harmonic emission IEC 6100-
3-2 
Class A
The X Series unit is suitable for use in all establishments, 
including domestic establishments and those directly 
connected to the public low-voltage power supply network 
that supplies buildings used for domestic purposes. 
Voltage fluctuations/ flicker 
emissions IEC 61000-3-3
Complies
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put 
into service according to EMC information provided in this document. 