beautypg.com

ZOLL X Series Monitor Defibrillator Rev H User Manual

Page 315

background image

Electromagnetic Compatibility Guidance and Manufacturer’s Declaration

9650-001355-01 Rev. H

X Series Operator’s Guide

A-29

Electromagnetic Compatibility Guidance and
Manufacturer’s Declaration

In-Flight Use (RTCA/DO-160):

The X Series unit complies with RTCA/DO-160, Environmental Conditions and Test
Procedures for Airborne Equipment, using the methods in Section 21, Category M for Radiated
and Conducted Radio Frequency Energy.

Guidance and manufacturer’s declaration – electromagnetic emissions

The X Series unit is intended for use in the electromagnetic environment specified below. The customer or the
user of the X Series unit should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment – guidance

RF Emissions CISPR 11

Group 1

The X Series unit uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.

RF emissions CSPR 11

Class B

Harmonic emission IEC 6100-
3-2

Class A

The X Series unit is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.

Voltage fluctuations/ flicker
emissions IEC 61000-3-3

Complies

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put
into service according to EMC information provided in this document.