Selecting 12-Lead Acquire
When enabled, the unit automatically prints the 12-Lead report when you press
. By default,
this feature is not enabled.
Specifying the Number of 12-Lead Print Copies
This option allows you to specify that the X Series unit print up to five copies of the 12-lead
waveform trace set after pressing
. By default, the unit prints only one 12-lead snapshot.
Specifying the 12-Lead Print Format
This option allows you to specify the print format for the 12-lead waveform traces: 3 x 4
(the default), 2 x 6, or 3 x 4 Cabrera format.
Printing 10 Seconds of the Lead ll Waveform Trace
This option allows you to specify that the unit print 10 seconds of the Lead
ll
waveform trace
after printing a 12-Lead snapshot.
Specifying the 12-Lead Frequency Response
This option allows you to specify the frequency response of the 12-lead waveform display.