Configuring NIBP Alarms and Settings

9650-001355-01 Rev. H

X Series Operator’s Guide

8-11

Setting Upper and Lower NIBP MAP Alarm Limits

Initially, the NIBP MAP Alarm Settings menu specifies that NIBP MAP alarms are enabled
(

ON

)

or disabled (

OFF

),

and displays the default Upper and Lower MAP limits. The upper and

lower limits can be

ON

or

OFF

(default is

OFF

). The following table lists the default MAP alarm

limits for adult, pediatric, and neonate patients, and gives the range in which you can set these
limits:

Using the NIBP Parameter Control Panel

Use the navigation keys to highlight and select the NIBP numeric display to display the NIBP
Parameter Control Panel
:

Figure 8-4 NIBP Parameter Control Menu

Patient Type

MAP Default

MAP Range

Adult

Lower: 50 mmHg

Upper: 120 mmHG

Lower: 25-230 mmHg

Upper: 23-228 mmHg

Pediatric

Lower: 50 mmHg

Upper: 110 mmHg

Lower: 25-140 mmHg

Upper: 23-138 mmHg

Neonate

Lower: 35 mmHg

Upper: 80 mmHg

Lower: 15-110 mmHg

Upper: 13-108 mmHg

NIBP

Lower

Upper

NIBP Systolic Alarm

75

220

NIBP Mode

Manual

NIBP Auto Mode Interval

5 min

NIBP Diastolic Alarm

35

110

NIBP MAP Alarm

50

120

Start TurboCuf

SmartCuf On/Off

On

59

50

100

150

200

250

300