beautypg.com

RIGOL DG5000 Series User Manual

Page 74

background image

RIGOL

Chapter 3 Arbitrary Waveform Output

User’s Guide for DG5000

3-14

automatically use straight lines to connect the start point (point 2, the level is
4 V) with the stop point (point 4, the level is -2 V) and automatically generate
point 1 and point 3 to create a continuous waveform.

3) After all the points are edited, press Return to return to the upper level menu,

then press Save to enter the file storage interface (refer to Store and
Recall
) and save the edited waveform.

7. View the edited waveform.

V

μs

4

2

0

-2

4

8

12

2

1

3

4


Note:
As can be seen from the above figure, the waveform editor inserts values
between the start point (point 2) and stop point (point 4) in the current interpolation
method and then automatically generates point 1 and point 3 to create a continuous
waveform.