RIGOL DG5000 Series User Manual
Page 51
Chapter 2 Basic Waveform Output
RIGOL
User’s Guide for DG5000
2-3
to input “500” and select the unit “mV
DC
” in the pop-up menu.
The offset range is limited by the “Resistance” and “Ampl/HoLevel”
setting. Please refer to the “Output Characteristic” described in
“Specifications”.
Amplitude and offset, HiLevel and LoLevel, always exist in pairs. Press the
softkey again to switch to “LoLevel”. Note that the low level should be at least
5 mV (50Ω) lower than the high level.
Selectable offset units: V
DC
, mV
DC
.
Selectable low level units: V, mV.
You can also use the knob to modify this parameter.
6. Set the Start Phase
Press Start Phase to highlight the softkey, and then use the numeric keyboard to
input “10” and select the unit “º” in the pop-up menu.
Start Phase range: 0 º to 360 º.
You can also use the knob to modify this parameter.
7. Channel Setting
Besides the above steps, you can also configure the output parameters related to
the channel through “Channel Setting” menu in Utility. As shown in the
following figure, the main output parameters (Resistance and Attenuation) of
the channel are shown at the bottom of the screen.
8. Enable the Output
Press Output of CH1 to highlight the key, which indicates the waveform output
from the [Output] connector of CH1 has been enabled. If an overload message
appears on the screen, please cut off the connection between the [Output]
connector and the external devices, and then press Output again to re-enable
the output.