Chapter 2 Basic Waveform Output

RIGOL

User’s Guide for DG5000

2-9

input “8” and select the unit “μsec” in the pop-up menu.
 Delay Time range: 0 s to pulse period
 Selectable Delay Time units: sec, msec, μsec, nsec.
 You can also use the knob to modify this parameter.

Restore Delay


Press Pulse, then use

to open the 2/2 menu page and press Restore, the

generator will align the delay between the two channels. Note that the restore delay
function is only for dual-channel models.

Assume that CH1 and CH2 output pulse waveforms with the same parameters. Use
the oscilloscope to sample the waveforms of the two channels and display them stably.
The two waveforms displayed on the oscilloscope have a certain delay. At this point,
press Restore on the generator and the waveforms will be displayed on the
oscilloscope without any delay.

Before Restore

CH1

CH2