Chapter 13 specifications, Chapter 13 specifications -1, Specifications – RIGOL DG5000 Series User Manual
Page 209
RIGOL
User’s Guide for DG5000
13-1
Chapter 13 Specifications
All the specifications can be guaranteed if the following two conditions are met unless
where noted.
The generator is within the calibration period and has performed self-calibration.
The generator has been working continuously for 30 minutes at specified
temperature (18°C ~ 28°C).
All the specifications are guaranteed unless those marked with “typical”.
Model
DG5352/
DG5351
DG5252/
DG5251
DG5102/
DG5101
DG5072/
DG5071
Channel
2/1
2/1
2/1
2/1
Maximum
Frequency
350 MHz
250 MHz
100 MHz
70 MHz
Sample Rate
1 GSa/s
Waveforms
Standard
Waveforms
Sine, Square, Ramp, Pulse, Noise
Arbitrary
Waveforms
Sinc, Exponential Rise, Exponential Fall, ECG, Gauss, HaverSine, Lorentz,
Dual-Tone, DC
Frequency Characteristics
Sine
1 μHz to 350 MHz 1 μHz to 250 MHz 1 μHz to 100 MHz 1 μHz to 70 MHz
Square
1 μHz to 120 MHz 1 μHz to 120 MHz 1 μHz to 100 MHz 1 μHz to 70 MHz
Ramp
1 μHz to 5 MHz
1 μHz to 5 MHz
1 μHz to 3 MHz
1 μHz to 3 MHz
Pulse
1 μHz to 50 MHz
1 μHz to 50 MHz
1 μHz to 50 MHz
1 μHz to 50 MHz
Noise
250 MHz Bandwidth
Arb
1 μHz to 50 MHz
1 μHz to 50 MHz
1 μHz to 50 MHz
1 μHz to 50 MHz
Resolution
1 μHz
Accuracy
±1 ppm, 18 °C to 28 °C
Sine Wave Spectrum Purity
Harmonic
Distortion
Typical (0 dBm)
≤100MHz:
Typical (0 dBm)
≤100MHz:
Typical (0 dBm)
≤100MHz:
Typical (0 dBm)
≤70MHz: