To Edit List


Press Edit ListEdit Point to open the

list editing interface.


1. Press Current Point, use the numeric keyboard or the knob to input the No. of

the point to be edited (i.e. n in the figure above), and this No. is limited by the
points of the current list.

2. Press Current Freq, use the numeric keyboard or the knob to input the

frequency of the current point (i.e. f

n

in the figure above), and this value is limited

by the maximum output frequency of the generator.

3. Press Return, finish point editing and return to the previous menu.

Press Save to open Store/Recall interface, save the edited FH

list. For detailed

introduction, please refer to

Store and Recall”.


To Create List


Press New List to open

list create interface.

1. Press List Points and use the numeric keyboard or the knob to input initial points.

The initial points could be up to 4096.

2. Press Edit Point to open the table edit interface.

 Press Current Point, use the numeric keyboard or the knob to input the No.

of the point to be edited, and this No. is limited by the current initial points.

 Press Current Freq, use the numeric keyboard or the knob to input the

frequency of the current point, and this value is limited by the maximum
output frequency of the generator.

 Press Return, finish point editing and return to the previous menu.

3. Press Save to open Store and Recall Interface, save the created FH

list. For

details, please refer to the introduction in

Store and Recall”.