Edit block, Edit block -13 – RIGOL DG5000 Series User Manual
Page 73
Chapter 3 Arbitrary Waveform Output
RIGOL
User’s Guide for DG5000
3-13
Edit Block
The following example introduces how to create an arbitrary waveform which meets
the following conditions with the following conditions met by using “Edit Block”.
Parameter
Value
Cycle Period
12 μs
High VLimit
4 V
Low VLimit
-2 V
Interpolation
Linear
Start
2, 4 V
End
4, -2 V
Operating Procedures:
1. Press Arb, use
to open the 2/2 menu page, and then press Create New
Cycle Period, use the numeric keyboard to input “12” and select the unit
“μsec” in the pop-up menu.
2. Press High VLimit, use the numeric keyboard to input “4” and select the unit “V”
in the pop-up menu.
3. Press Low VLimit, use the numeric keyboard to input “-2” and select the unit “V”
in the pop-up menu.
4. Press Init #Points, use the numeric keyboard to input “4” and then press OK.
Note that a level line of -2 V appears.
5. Press Interp to select “Linear”.
6. Press Edit Block to enter the “Edit Block” interface.
1) Press X1 and use the numeric keyboard to input “2”. Then press Y1 and use
the numeric keyboard to input “4” and select the unit “V” in the pop-up menu.
Use the same method to set X2 and Y2.
2) Press Excute to apply the setting defined in step 1). The waveform editor will