beautypg.com

RIGOL DG5000 Series User Manual

Page 34

background image

RIGOL

Chapter 1 Quick Start

User’s Guide for DG5000

1-10

26. CH1 Output

This BNC connector is used as an output terminal.
For dual-channel model: enable or disable waveform signals generated from
[Output] connector corresponding to CH1. The nominal output impedance is 50 Ω.
For single-channel model: output a TTL-compatible pulse synchronized with the
main output. The nominal source impedance is 50 Ω.

27. CH2 Output

This BNC connector is used as an output terminal. The nominal output impedance
is 50 Ω.
For dual-channel model: enable or disable waveform signals generated from
[Output] connector corresponding to CH2.
For single-channel model: output signals of the main channel.

CAUTION

Overvoltage protection of the output channel will take effect once any of
the following conditions is met.
 Amplitude setting in the generator is greater than 2 Vpp; the input

voltage is greater than ±12.1 V (±0.1 V) and frequency is lower than
10 kHz.

 Amplitude setting in the generator is lower than or equal to 2 Vpp; the

input voltage is greater than ±4.8 V (±0.1 V) and frequency is lower
than 10 kHz.

The message “OverLoad protect, The output is off!” will appear on the
screen when overvoltage protection takes effect.