beautypg.com

Sync polarity, Output polarity, Sync polarity -5 output polarity -5 – RIGOL DG5000 Series User Manual

Page 155

background image

Chapter 10 Advanced Operation

RIGOL

User’s Guide for DG5000

10-5

Sync Polarity


Set the sync signal on the Sync connector as normal or invert.

Press UtilityCH1 SetPolarity to select “Positive” or “Negative”. Notice that
the sync signal related to the waveform is not invert even when the waveform is
invert.

Output Polarity


Set the signal on the [Output] connector as normal or invert. Waveform inverts
relatively to the offset voltage.

Press UtilityCH1 SetOutput to select “Normal” or “Invert”. The default is
“Normal”. The first half period of the cycle waveform is positive in “Normal” mode and
negative in “Invert” mode.

0 V

Offset

0 V

Normal Invert

Normal Invert


After the waveform is inverted, please pay attention to the following cases:
 None offset voltage is changed.
 The waveform viewed in Graphic mode is not invert.
 The sync signal related to the waveform is not invert.